1. Вступ
The Extech DO600 Waterproof ExStik II Dissolved Oxygen Meter is designed for precise measurement of dissolved oxygen levels. It displays oxygen concentration as a percentage of saturation (0 to 200%) or as a concentration in parts per million (ppm) or milligrams per liter (mg/L) (0 to 20.00 ppm/mg/L). Key features include adjustable altitude compensation, adjustable salinity compensation, and an analog bargraph for trend indication. This manual provides instructions for the proper setup, operation, and maintenance of your DO600 meter.
2. Вміст упаковки
Переконайтеся, що всі предмети є в наявності та у належному стані після розпакування:
- Extech DO600 Dissolved Oxygen Meter
- Розчинений електрод
- Protective Sensor Cap
- Spare Membrane Cap
- Electrolyte Solution
- Four 3V CR-2032 Batteries (pre-installed or included separately)
- 48-inch (1.2m) Neck Strap

Figure 1: Extech DO600 meter with its accessories, including the electrode, caps, electrolyte, and neck strap.
3. Компоненти лічильника

Малюнок 2: Спереду view of the Extech DO600 meter.
- РК-дисплей: Shows dissolved oxygen readings, temperature, and other indicators.
- Кнопка РЕЖИМ/УТРИМАННЯ: Toggles between measurement modes and holds the current reading.
- CAL/RECALL Button: Initiates calibration and recalls stored data.
- Кнопка ON/OFF: Powers the meter on or off.
- електрод: Sensor for dissolved oxygen measurement.
- Protective Sensor Cap: Covers and protects the electrode.
4. Налаштування
4.1. Встановлення батареї
The DO600 meter uses four 3V CR-2032 batteries. These are typically pre-installed. If replacement is necessary:
- Twist off the battery compartment cap located at the top of the meter.
- Remove the old batteries, noting their polarity.
- Insert new CR-2032 batteries with the correct polarity.
- Securely replace the battery compartment cap to maintain the waterproof seal.
4.2. Electrode Preparation
Before first use or after prolonged storage, the DO electrode requires preparation:
- Unscrew the protective sensor cap from the electrode.
- Fill the spare membrane cap with the provided electrolyte solution. Ensure no air bubbles are trapped.
- Carefully screw the membrane cap onto the electrode, ensuring the membrane is taut and free of wrinkles.
- Allow the electrode to stabilize for approximately 15-30 minutes in the electrolyte solution before calibration.
5. Інструкція з експлуатації
5.1. Увімкнення/вимкнення живлення
Натисніть УВІМК./ВИМК button to power the meter on or off.
5.2. Калібрування
Regular calibration ensures accurate readings. Calibrate the meter in air-saturated water or a known standard solution.
- Ensure the electrode is prepared as described in Section 4.2.
- Immerse the electrode in air-saturated water (e.g., tap water stirred vigorously for several minutes) or a DO standard solution.
- Натисніть і утримуйте CAL/Відкликання кнопку, доки на дисплеї не з’явиться напис «CAL».
- The meter will automatically recognize the calibration point and adjust. Wait for the reading to stabilize and for the calibration indicator to confirm completion.
5.3. Зняття мірок
To measure dissolved oxygen:
- Увімкніть лічильник.
- Занурте електрод уample solution. Ensure the membrane is fully submerged and free of air bubbles.
- Обережно перемішайте електрод уample to ensure proper flow across the membrane.
- Wait for the reading on the LCD display to stabilize. The display will show both dissolved oxygen and temperature.
- Натисніть РЕЖИМ/УТРИМКА button to cycle between % saturation and ppm (mg/L) readings.
- To hold a reading, briefly press the РЕЖИМ/УТРИМКА кнопку. Натисніть ще раз, щоб відпустити.

Figure 3: The Extech DO600 meter being used to measure dissolved oxygen in a laboratory.
5.4. Altitude Compensation
Dissolved oxygen saturation levels are affected by altitude. The DO600 allows for manual adjustment:
- Refer to the meter's full instruction manual for specific steps to enter and adjust altitude compensation (0 to 20,000 feet in 1,000 ft. increments).
5.5. Salinity Compensation
Salinity also influences dissolved oxygen levels. The DO600 provides adjustable salinity compensation:
- Consult the meter's comprehensive manual for detailed instructions on setting salinity compensation (0 to 50 ppt).
6. Технічне обслуговування
6.1. Electrode Cleaning and Storage
To maintain electrode performance:
- Промивайте електрод дистильованою водою після кожного використання.
- If the membrane becomes fouled, gently clean it with a soft cloth or cotton swab. Replace the membrane cap if it is damaged or heavily contaminated.
- When storing, ensure the protective sensor cap is filled with electrolyte solution to keep the membrane hydrated.
6.2. Заміна батареї
Замініть батарейки, коли на дисплеї з’явиться індикатор низького заряду батарейок. Інструкції щодо встановлення батарейок дивіться у розділі 4.1.
7. Вирішення проблем
| проблема | Можлива причина | Рішення |
|---|---|---|
| Лічильник не вмикається | Розряджені батарейки; неправильне встановлення батарейок | Замініть батарейки; перевірте полярність батарейок |
| Неточні або нестабільні показники | Electrode not calibrated; Fouled membrane; Air bubbles on membrane; Insufficient electrolyte | Perform calibration; Clean/replace membrane; Remove air bubbles; Refill electrolyte |
| No reading displayed | Electrode not connected properly; Damaged electrode | Ensure electrode is securely attached; Replace electrode if damaged |
8. Технічні характеристики
| Параметр | Значення |
|---|---|
| Модель | DO600 |
| Dissolved Oxygen Range (Concentration) | від 0 до 20.00 ppm (мг/л) |
| Dissolved Oxygen Range (Saturation) | від 0 до 200.0 % |
| Діапазон температур | Not explicitly stated for display, but upper rating is 50 ℃ |
| Компенсація висоти | 0 to 20,000 feet (in 1,000 ft. increments) |
| Компенсація солоності | Від 0 до 50 ppt |
| Джерело живлення | Four 3V CR-2032 batteries |
| Розміри | 6.9 x 1.6 x 1.4 дюйми (17.5 x 4.1 x 3.6 см) |
| Вага товару | 3.87 унцій (0.11 кг) |
| Рейтинг водонепроникності | Yes (implied by "Waterproof ExStik II") |

Figure 4: Physical dimensions of the Extech DO600 meter. Note: The image shows 7.7 inches, while product specifications state 6.9 inches. Refer to the latest product documentation for precise measurements.
9. Відео про продукти
У наданих даних не знайдено офіційних відео від продавця для цього товару.
10. Гарантія та підтримка
For warranty information, technical support, or to purchase accessories and replacement parts, please refer to Extech's official website or contact their customer service department. You can find more information on the Extech Store on Amazon.